• Waterproof (ip66) housing
• Auto power off function datalog up to 100 points
• Low battery alarm baclit custom LCD display
• Rechargeable battery
• Certificated CE mark GD mark , ISO 9001
• Short response time and easy-to-use using
high sensitive polarograpich type electrode internal diagnosis
for normal operation check B1-directional RS232C allows remote
operation via Pc Simultaneously displays DO (or 02/air/p02) temp
and time automatic temp and barometric pressure compensation.
♧ Specification
DO
• Range : 0.00 ~ 19.99mg / ℓ
• Resolution : 0.1 / 0.01
• Relative Accuracy : ± 0.5%
O2
• Range : 0.0 ~ 60.0%
• Resolution : 0.1%
• Relative Accuracy : ± 1 digit
Air Saturation (%)
• Range : 0.0 ~ 199.9%
• Resolution : 0.1%
• Relative Accuracy : ± 1 digit
Temperature
• Range : -10 ~60℃
• Resolution : 0.1℃
• Relative Accuracy : ± 0.4℃
Dimensions / Weight
• 220(L) X 80(W) X 40(H) / 450g(Body)
Features and specifications
1. PH-20N
Feature 1-1
• Large and clear Custom LCD
• Design of waterproof design structures
• Battery remaining indicator
• Rechargeable batteries and adapters are included
• Installation support that can be used in the laboratory
• Simultaneous display of measurement data and temperature
and current time display
• Built-in portable case and calibration solution
• Automatic calibration of standard buffers
• Measurement of measurement stability
• Acquired ISO9001, GD, CE mark
1-2 Specifications
PH
• Measuring range: -2.00 to 19.99
• Resolution: 0.01 / 0.1
• Accuracy: ± 0.02
Temperature
• Measuring range: -10 to 110 ° C
• Resolution: 0.1 ° C
• Accuracy: ± 0.4 ℃
Millivolts
• Measuring range: 1999.9mV
• Resolution: 0.1 mV
• Accuracy: ± 0.1mV
Product Size
220 (L) X 80 (W) X 40 (H) / 450 g (Agency)
2. PH-25N
Feature 2-1
• Large and clear Custom LCD
• Design of waterproof design structures
• Battery remaining indicator
• Rechargeable batteries and adapters are included
• Installation support that can be used in the laboratory
• Simultaneous display of measurement data and temperature
and current time display
• Built-in portable case and calibration solution
• Automatic calibration of standard buffers
• Measurement of measurement stability
• Acquired ISO9001, GD, CE mark
• Show the ion concentration
2-2 Specifications
PH
• Measuring range: -2,000 to 19,999
• Resolution: 0.001 / 0.1
• Accuracy: ± 0.002
Concentration
• Measuring range: 0,0001 to 19,999 mg
• Resolution: 0.001
• Accuracy: ± 0.25% of Reading
Temperature
• Measuring range: -10 to 110 ° C
• Resolution: 0.1 ° C
• Accuracy: ± 0.4 ℃
Millivolts
• Measuring range: 1999.9mV
• Resolution: 0.1 mV
• Accuracy: ± 0.1mV
♧ Feature
• 4,7,10 PH automatic recognition a buffer
in automated and manual calibration
• wall or panel - mounted housing
• Automatic and manual temperature compensation
• Easy operation by front keys
• Output current, temperature, measured values and a token value emf
• User - defined alarm configuration
• Automatic cleaning function (with chemoclean) trigggered
by alarm or limit switch
♧ Aplication
• For drinking water and waste water treatment processes
• For measuring plate production lines of production
and processing equipment of water
• Food & drug process controls
• For Measuring quality purifying plants, sewage
and waste water treatment plant
• For measuring quality of aqua culture - and fishery
• For power plant control and swimming pool water
♧ Technical Specifications for Per Model
1.Transmitter
● Measuring Principle Glass electrode
● Resolution 0.001 pH
● Measuring Range 0 ~ 14.0 pH / -2000 ~ +2000mV
● Accuracy ±0.1% of Full Scale Span
● Repeatability 0.02pH / 1mV of Full Scale
● Response Time 15sec (90% Saturation)
● Operating Termperature -30~80°C
● Temperature Compensated 0~50°C (Auto or Manual Compensated)
● Relative Humidity 0 to 95%, None-Condensing
● Analog output 4-20 mA
● Communication output RS232C (OPTION)
● Alarm Contact High, Low, Clean 250VAC 0.5A SPST
● Load Resistance 750ohm Max. Load
● Power Voltage AC110 ~ 220V 50/60Hz ±10%, approx. 3Watt
2.Model
1.CUBE-PH1000
• Display : 3.5 Inch Graphic LCD
• Output : 2 Chanel Isolated 4~20m ADC
• Materials : Glass Fiber Nylon
• Enclosure Rating : NEMA 4X
• Dimension : 96X96X155
• Install : Panel Mount
2.CUBE-PH2000
• Display : 5.3 Inch Graphic LCD
• Output : 2 Chanel Isolated 4~20m ADC
• Materials : Glass Fiber Nylon
• Enclosure Rating : NEMA 4X / IP 65
• Dimension : 144X144X185
• Install : Panel or Wall / Stanchion Mount